【摘要】 :[目的] 观察YKX/Y型医用空气消毒机(以下简称消毒机)除菌净化的效果 。[方法] 用消毒机进行空气消毒,然后用自然沉降法监测消毒结果,于开机消毒后2h、3h、4h分别在3个手术间4角和中央放5个平皿,打开暴露5min然后置37℃温箱培养48h做菌落记数,并与传统的紫外线空气消毒方法比较。[结果] 两组空气消毒后各时限平均菌落数比较,差异有显著性(P<0.01)。该消毒机进行空气消毒效果优于传统的紫外线空气消毒法。[结论] 该消毒机适用于手术室空气的持续消毒,是目前比较满意的一种新型空气消毒方法。
【关键词】 空气消毒机;净化;紫外线消毒
Abstract:[Objective] To observe sterilization purification effect of YKX/Y type medical air disinfection machine.[Method] Make air disinfection with the disinfection machine,and then supervise the disinfection result with natural sedimentation,put 5 flat utensils in 4 angles and center of 3 operation rooms after 2h,3h,4h respectively,open and expose for 5min,then put in warm box of 37℃ for 48h,make colony count,compare it with traditional ultraviolet radiation air disinfection.[Result] Comparing average colony count of every time point after disinfection,the difference is marked(P<0.01).The machine is better than the traditional one.[Conclusion] The disinfection machine is fit for continual disinfection of operation room,and is presently satisfying new air disinfection.
Key words:air disinfection machine;purification;ultraviolet radiation disinfection
空气消毒是控制医院感染的重要环节。如何进行有效的空气消毒,减少空气污染造成的医院感染,特别对高危易感科室的空气持续消毒并保持洁净度[1],近年来尤为人们所关注。我院自2006年1月开始在手术室安装了消毒机,通过监测并与传统的紫外线消毒方法进行对比,效果满意,现报道如下。
1 材料与方法
1.1 材料
选择普通手术间6间,其中50m3的手术间4间,40m3的手术间2间。分别设观察组和对照组,平均分配手术室。两组实验条件均衡,无统计学差异(P>0.05),具有可比性。YKX/Y型医用空气消毒机,医用紫外线灯,37℃温箱,平皿若干。
1.2 方法
观察组:用肯格王牌YKX/Y型医用消毒机(中英合资成都肯格王有限公司生产)两台分别置于手术台两侧,按要求高档换气,分别对手术间进行消毒。按卫生部《消毒技术规范(2002年版)》中相关要求进行,于消毒后2h、3h、4h采用平皿沉降法,在手术间4角和中央共取5点,4角均距离墙壁1m,采样高度距地面1.5m。采样时将平板盖打开,扣放于平板旁,暴露5 min,盖好立即送检。将送检的平板置37℃温箱培养48h,按平板暴露法计算公式CFU/m3=50 000N/AT计算菌落数。根据国家有关医院卫生消毒标准手术室空气细菌平均菌落数要求200cfu/m3为合格。对照组:手术室内未放置消毒机,行紫外线消毒,其余同观察组。 1.3 统计学方法 [1] [2] 下一页 |